Laboratoric thermostatic chamber version (temperature only); 112 liters; [-40°C; +180°C]
GENERAL INFORMATION
– Useful capacity: 112 l
– Internal dimensions (WxDxH): 548x460x447 mm
– Ext. dimensions (WxDxH): 880x1065x1380 mm (with wheels table: 904x1120x1937)
– Rated power: 3 kW
– Rated current absorption: 19A
– Voltage: 230 V ±10%/50Hz/1ph + G
– Noise: 58 dBA (measured at 1 m from the front wall in a non reverberating room)
– Weight: 385 kg
PERFORMANCES
(measured at room temperature of 22°C and empty working space)
– Temperature range: [-40°C; +180°C]
according to IEC 60068-3-5, IEC 60068-3-6
– Temperature Fluctuation: ±0,1°C... ±0,3°C
– Temperature Rate of Change (measured at the control probe):
Heating = 3.2°C/min. from -40°C to +180°C
Cooling = 4°C/min. from +180°C to -40°C
– Thermal Load at -25°C: 100 W
CONTROLS
controlled by PLC and "KeyKratos Plus" touchscreen on the door.
Memory capacity to 1000 cycles with 350 segments.
Unlimited real time recording of temperature and humidity versus time.
Recordings in CSV format (Comma Separated Value) for easy export to Excel® from Memory Card to Pendrive via USB interface placed on the front panel (program files are easily convertible into graphic format).
A trend chart for recorded variables available with a scale from -100/+200.
Messages on scheduled maintenance activities.
Synoptic graphs for the chamber, humidity, low stage refrigeration and high stage refrigeration.
THE BASIC CONFIGURATION INCLUDES:
* Mechanical door closers
* 300x300 mm multipane inspection window
* n° 1 shelf
* n° 1 lateral porthole (diam. 80 mm) on the right side
* lnternal lighting
* Air condenser
* RS232 interface
* Ethernet port
* Digital undertemperature/overtemperature switch with indipendent probe
* T analogic retransmission
* no. 1 auxiliary contact (specimens)
* Connection to UPS
The software source code and the relative documentation are not included.
DY110T ACCESSORIES
Code | Name |
18AX0007 | CE MARK (INCLUDING EMC) (for 110 l) |
18AX0016 | SUPPORTING TABLE WITH WHEELS (for 110 l) chamber fixed on the table, overall dims: mm 904x1120x1937 (LxDxH) |
18AX0025 | SET OF NO. 4 ANALOGIC INPUTS 0÷10V FOR USER'S DATA ACQUISITION (no. 1 set max.) |
18AX0026 | SET OF NO. 4 PT100 INPUTS (no. 1 set max.) |
18AX0027 | SET OF NO. 8 AUXILIARY CONTACTS (no. 1 set max.) (16, 60 l excluded) |
18AX0074 | ULTRAVIOLET RAY LAMP (for 110 l) |
18AX0089 | INTERNAL SHELF 110 l |
18AX0111 | 80 MM PORTHOLE >=110 l |
18AX0123 | OVERPRICE FOR SPECIAL VOLTAGES ON REQUEST |
18AX0140 | WINKRATOS S/W (WINDOWS) FOR ONE CHAMBER REMOTE CONTROL BY PC (via Ethernet) |
18AX0141 | P.C. COMPLETE WITH WINKRATOS S/W (WINDOWS) FOR ONE CHAMBER REMOTE CONTROL (via Ethernet) |
18AX0142 | WINKRATOS S/W (WINDOWS) FOR MULTICHAMBER CHAMBER REMOTE CONTROL BY PC (via Ethernet, max 16 chambers) |
18AX0146 | P.C. COMPLETE WITH WINKRATOS S/W (Windows) FOR MULTICHAMBER REMOTE CONTROL (via Ethernet, max 16 chambers) |
18AX0151 | WINKRATOS S/W (WINDOWS) FOR MULTICHAMBER CHAMBER REMOTE CONTROL BY PC (via RS232 or RS485 or RS422, max 16 chambers) ON REQUEST |
18AX1290 | NO BREAK POWER UNIT FOR PLC 110 l |
18AX1291 | SPECIMEN SWITCHING OFF IN CASE OF CHAMBER ALARM 110 l |
19AX0180 | CALIBRATION CERTIFICATE FOR TEMPERATURE (USING INSTRUMENTS CALIBRATED WITH ACCREDITED CALIBRATION LABORATORY METROLOGICAL CHAIN) |