Discovery Climatic Chamber DY340C

This version of climatic test chambers can control temperature [-75°C; +180°C] and Relative Humidity [10% to 98%], volume 336 liters;

GENERAL INFORMATION

 – Useful capacity: 336 l

 – Internal dimensions (WxDxH): 601X810X692 mm

 – Ext. dimensions (WxDxH): 919X1817X1638* mm (*height without wheels = decrease of 60 mm)

 – Electrical power: 9 kW (average 5.2 kW)

 – Voltage: 400 V ±10%/50Hz/3ph + N + G

 – Noise: 63 dBA (measured at 1 m from the front wall in a non reverberating room)

 – Weight: 730 kg

 PERFORMANCES

 (measured at room temperature of 22°C and empty working space)

 – Temperature range: [-75°C; +180°C]

 according to IEC 60068-3-5, IEC 60068-3-6

 – Relative Humidity range: from 10% to 98% within the temperature range [+5°C; +95°C]

 – Dew Point range: from +2°C to +94°C for continuous tests; down to -20°C for limited periods based on starting conditions and foreseen testing time

 – Temperature Fluctuation: ±0,1°C... ±0,3°C

 – Temperature Rate of Change (measured at the control probe):

 Heating = 4.5°C/min. from -70°C to +180°C

 Cooling = 2°C/min. from +180°C to -70°C

 – Thermal Load at -55°C: 600 W

 – Relative Humidity Fluctuation ±1%...±3%

 CONTROLS

 controlled by PLC and "KeyKratos Plus" touchscreen on the door.

 Memory capacity to 1000 cycles with 350 segments.

 Unlimited real time recording of temperature and humidity versus time.

  Recordings in CSV format (Comma Separated Value) for easy export to Excel® from Memory Card to Pendrive via USB interface placed on the front panel (program files are easily convertible into graphic format).

 A trend chart for recorded variables available with a scale from -100/+200.

 Messages on scheduled maintenance activities.

 Synoptic graphs for the chamber, humidity, low stage refrigeration and high stage refrigeration.

 THE BASIC CONFIGURATION INCLUDES:

 * "Gatekeeper®" electromagnetic door closers

 * Self-pivoting wheels

 * 450x450 mm multipane inspection window

* Self feeding system

 * n° 1 shelf

 * no. 1 lateral porthole (diam. 80) on the left side

 * no. 1 lateral porthole (diam. 150) on the right side

 * lnternal lighting

 * Air condenser

* RH measurement by psychrometric system

 * RS232 interface

 * Ethernet port

 * Digital undertemperature/overtemperature switch with indipendent probe

 * T & RH analogic retransmission

 * no. 1 auxiliary contact (specimens)

 * Connection to UPS

 The software source code and the relative documentation are not included.

 * MyAngel24™ remote assistance system, complete with a SIM card for wireless connection via GPRS. On request LAN connection via Ethernet can be supplied. MyANgel24 activation is subject to customer’s confirmation by attached questionnaire to be completed and returned with the order.

 The software source code and the relative documentation are not included.

 PAPILDOMAS PASIRINKIMAS DY340 C

 Code Name
18AX0003 CE MARK (INCLUDING EMC)
18AX0025 SET OF NO. 4 ANALOGIC INPUTS 0÷10V FOR USER'S DATA ACQUISITION (no. 1 set max.)
18AX0026 SET OF NO. 4 PT100 INPUTS (no. 1 set max.)
18AX0027 SET OF NO. 8 AUXILIARY CONTACTS (no. 1 set max.) (16, 60 l excluded)
18AX0072 ULTRAVIOLET RAY LAMP (for 250 - 340 l)
18AX0100 INTERNAL SHELF 340 l
18AX0108 150 MM PORTHOLE >=250 l
18AX0111 150 MM PORTHOLE >=250 l
18AX0116 NOISE INSULATION 250 - 340 l (water condensation, not available for 250 l ES version). Add 20 working days to the delivery term.
18AX0120 OVERPRICE FOR 380/50/3 WITHOUT N >=250 l
18AX0123 OVERPRICE FOR SPECIAL VOLTAGES ON REQUEST
18AX0140 WINKRATOS S/W (WINDOWS) FOR ONE CHAMBER REMOTE CONTROL BY PC (via Ethernet)
18AX0141 P.C. COMPLETE WITH WINKRATOS S/W (WINDOWS) FOR ONE CHAMBER REMOTE CONTROL (via Ethernet)
18AX0142 WINKRATOS S/W (WINDOWS) FOR MULTICHAMBER CHAMBER REMOTE CONTROL BY PC (via Ethernet, max 16 chambers)
18AX0145 NO BREAK POWER UNIT FOR PLC >=250 l
18AX0146 P.C. COMPLETE WITH WINKRATOS S/W (Windows) FOR MULTICHAMBER REMOTE CONTROL (via Ethernet, max 16 chambers)
18AX0151 WINKRATOS S/W (WINDOWS) FOR MULTICHAMBER CHAMBER REMOTE CONTROL BY PC (via RS232 or RS485 or RS422, max 16 chambers) ON REQUEST
18AX0170 SPECIMEN SWITCHING OFF IN CASE OF CHAMBER ALARM >=250 l
19AX0181 CALIBRATION CERTIFICATE FOR TEMPERATURE AND R.H. (USING INSTRUMENTS CALIBRATED WITH ACCREDITED CALIBRATION LABORATORY METROLOGICAL CHAIN)